Description
This TMA was specifically developed
to measure the CTEs of Polymers and Resins
in the temperature range -30 to +70°C.
The instrument is built into
a very compact nicely designed housing, where
all necessary components are included.
The sample can be inserted after the protecting
cover is opened from the top of the instrument.
The sample is loaded between a fixed point
and a moveable push rod out of quartz glass,
which is connected to a high precision LVDT
sensor, with 10nm / 1,25nm/digit resolution.
Sample dimensions of maximum 50 mm length
and 8, 5 mm diameter are possible. An option
is available, so that the hardening processes
of Resins can be measured.
Technical Specifications
| |
TMA - PT10 |
| Temperature Range |
-30 to +70°C |
| Max. sample size |
50 mm (L) , 8,5 mm (D) |
| Measurement precision |
+/- 0.1% |
| Resolution |
0.2nm |
| RMS Noise |
3.5nm |
| Dynamic baseline drift |
<1µm |
| Force range |
0.001 to 1 N |
| Force resolution |
0.001 N |
| Mass Flow Control |
Optional |
| Atmosphere |
Air |
| Operation Modes |
|
| Standard |
Included |
| Optional |
|
| DTA feature |
Optional |
|