Description

The LINSEIS STA PT1600 HP (simultaneous thermal analysis) can be used to determine simultaneous changes of mass Thermogravimetry (TG) and caloric reactions Differential Scanning Calorimeter (DSC) of a sample in the temperature range RT up to 1750°C. Furthermore the entire system can withstand a pressure of up to 2 or 5 bar at any temperature. The pressure is controlled through a unique pressure control system.

The system enables Thermo gravimetric (TG) or Simultaneous Thermal Analysis (TG-DTA/DSC) measurements under pressure, reducing, inert, oxidizing and vacuum atmospheres. The system can measure in steady or dynamic atmospheres. The unique pressure control system allows dynamic pressure adjustments during the entire measurement. Software controlled MFC´s (Mass Flow Controllers) are certainly also available.

Model

 

STA PT1600 HP

Temperature range

 

RT up to 1000°C
RT up to 1750°C

Pressure at 1000/1750°C
Vacuum:

 

2 or 5 bar (two versions)
10E-
5 mbar (with Turbomolecular Pump)

TGA

 

 

Sample Size

 

up to 25g

Resolution

 

0.5 µg

RMS noise

 

< 1µg

TG-DSC

 

 

HDSC resolution

 

0.3, 0.4, 1µW

HDSC RMS noise

 

4, 6, 17.5µW

HDSC sensor type

 

E, K, S, B

TG-DTA

 

 

DTA Sensitivity

 

0.05µV

Special Features

 

Pressure control system
Mass flow controllers for up to four gases
Back pressure regulator